• Circular sterilisation chamber with square or rectangular hatches (doors).
  • The TS-AS model boasts a flat/circular cavity that completely envelopes the chamber to quickly and uniformly heat/cool the loads.
  • The TS-OW model is single-walled (no cavity).
  • Some reinforcements are welded onto the outside of the chamber and/or cavity to ensure stability during operating conditions..
  • The chamber is positioned on a base that has been designed specifically to evenly distribute the weight on the floor.
  • The loading trolley-inserting and removing guides are positioned on the bottom of the chamber; they are also height-adjustable.
  • The bottom of the chamber converges towards the drain sump in order to make disposing of the condensate easier.
  • Two 2” tri-clamp ferrules are welded to the chamber for validation purposes (temperature uniformity mapping and pressure tests).
  • All of the chambers internal surfaces are mechanically polished and then subjected to chemical degreasing, pickling and passivation treatments. Surface finish: ≤ 0.35 micron.
  • Above-floor or recessed chamber installation possible (depending on the size).