- Rectangular sterilisation/depyrogenation chamber and hatches (doors).
- Fully welded one-piece chamber with rounded corners to ensure excellent hygiene levels and facilitate internal surface cleaning.
- The chamber/base is fitted with a cavity to quickly and uniformly heat and cool loads.
- The mechanical construction is resistant to temperatures above 350°C.
- Air is uniformly distributed inside the chamber by means of perforated walls with multi-diameter holes (holes with a bigger diameter towards the bottom and holes with a smaller diameter towards the top).
- The air recirculation and distribution system is installed in the upper part of the steriliser. The heating elements and cooling exchanger are installed concentrically with respect to the impeller of the air recirculation fan.
- The chamber is positioned on a tubular-type base to evenly distribute the weight on the floor.
- The loading trolley-inserting and removing guides are positioned on the bottom of the chamber; they are also height-adjustable.
- Two 1.5” tri-clamp ferrules are welded to the chamber for validation purposes (air particle sampling and temperature uniformity mapping).
- All of the chambers internal surfaces are mechanically polished and then subjected to chemical degreasing, pickling and passivation treatments. Surface finish: ≤ 0.35 micron.
- Above-floor chamber installation.
- Both doors (loading and pick-up side) have been designed to be connected to the walls of the clean chambers (“Bio-seal” interface).